Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Tek DM251: (1 year) CAL VER /5520 DATE: 26-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 52 NUMBER OF LINES: 189 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke Corp. makes no warranty, expressed or implied, as to the fitness or # useability of this procedure. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: NONE 1.004 DISP Temperature: 18 - 28 degC 1.004 DISP Humidity: <80% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD INITIAL SETUP 1.007 DISP Connect the 5520A to the UUT as follows: 1.007 DISP [32] 5520A NORMAL HI to UUT V OHM 1.007 DISP [32] 5520A NORMAL LO to UUT COM 1.008 HEAD {AC VOLTAGE TESTS} 1.009 DISP Set the UUT to ACV. 1.010 RNG 4V AC 1.011 ASK- R 1.012 HEAD AC VOLTAGE TESTS: {4V Range} 1.013 5520 4 0V 0.005U 2W 2.001 5520 4 2.000V 0.019U 50H SI 2W 3.001 5520 4 2.000V 0.025U 1kH SI 2W 4.001 ASK+ R 4.002 HEAD AC VOLTAGE TESTS: {40V Range} 4.003 5520 40 20.00V 0.19U 50H SI 2W 5.001 5520 40 20.00V 0.25U 1kH SI 2W 6.001 HEAD AC VOLTAGE TESTS: {400V Range} 6.002 5520 400 200.0V 1.9U 50H SI 2W 7.001 5520 400 200.0V 2.5U 1kH SI 2W 8.001 HEAD AC VOLTAGE TESTS: {750V Range} 8.002 5520 750 750V 10U 50H SI 2W 9.001 5520 750 750V 13U 1kH SI 2W 10.001 HEAD {DC VOLTAGE TESTS} 10.002 DISP Set the UUT to DCV. 10.003 HEAD DC VOLTAGE TESTS: {400mV Range} 10.004 5520 400 0.0mV 0.2U 2W 11.001 5520 400 300.0mV 1.1U 2W 12.001 5520 400 -300.0mV 1.1U 2W 13.001 5520 400 200mV 0.8U 2W 14.001 HEAD DC VOLTAGE TESTS: {4V Range} 14.002 5520 4 2.000V 0.008U 2W 15.001 HEAD DC VOLTAGE TESTS: {40V Range} 15.002 5520 40 20.00V 0.08U 2W 16.001 HEAD DC VOLTAGE TESTS: {400V Range} 16.002 5520 400 200.0V 0.8U 2W 17.001 HEAD DC VOLTAGE TESTS: {1000V Range} 17.002 5520 1000 1000V 5U 2W 18.001 HEAD {RESISTANCE TESTS} 18.002 DISP Set the UUT to OHMS. 18.002 DISP 18.002 DISP Press the BLUE button to select RESISTANCE. 18.003 HEAD RESISTANCE TESTS: {40 MOhm Range} 18.004 5520 40 19.00MZ 0.62U 2W 19.001 HEAD RESISTANCE TESTS: {400 kOhm Range} 19.002 5520 400 190.0kZ 1.0U 2W 20.001 HEAD RESISTANCE TESTS: {4000 kOhm Range} 20.002 5520 4000 1900kZ 15U 2W 21.001 HEAD RESISTANCE TESTS: {40 kOhm Range} 21.002 DISP Connect the 5520A to the UUT as follows: 21.002 DISP [32] 5520A SENSE HI to UUT V-OHM 21.002 DISP [32] 5520A SENSE LO to UUT COM 21.002 DISP [32] 5520A NORMAL HI to UUT V-OHM 21.002 DISP [32] 5520A NORMAL LO to UUT COM 21.002 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 21.002 DISP the NORMAL leads to the SENSE leads AT the UUT. 21.003 5520 40 19.00kZ 0.10U CW 22.001 HEAD RESISTANCE TESTS: {4 kOhm Range} 22.002 5520 4 1.900kZ 0.010U CW 23.001 HEAD RESISTANCE TESTS: {400 Ohm Range} 23.002 5520 400 190.0Z 1.2U CW 24.001 5520 0.0Z +0.4U CW 25.001 HEAD {CONTINUITY TEST} 25.002 DISP Press the BLUE button to select CONTINUITY. 25.003 5520 0.0Z S CW 25.004 EVAL -s "Audible Tone" : Is there an Audible Tone? 26.001 HEAD {DIODE TESTS} 26.002 DISP Connect the 5520A to the UUT as follows: 26.002 DISP [32] 5520A NORMAL HI to UUT V OHM 26.002 DISP [32] 5520A NORMAL LO to UUT COM 26.003 DISP Set the UUT function to DIODE/CAPACITANCE. 26.003 DISP 26.003 DISP Press the BLUE button to select DIODE. 26.004 MATH S[1] = "UUT indicates between 2.000 and 3.600" 26.005 EVAL -s S[1] : Does UUT indicate between 2.000 and 3.600? 27.001 ASK- U # Insure that lead resistance is less than 10 Ohms. 27.002 5520 0.0Z S 2W 27.003 ASK+ U 27.004 MATH S[1] = "UUT indicates between -0.001 and 0.001" 27.005 EVAL -s S[1] : Does UUT indicate between -0.001 and 0.001? 28.001 HEAD {CAPACITANCE TESTS} 28.002 DISP Press the BLUE button to select CAPACITANCE. 28.003 5520 2.000nF 0.06U 2W 29.001 5520 20.00nF 0.24U 2W 30.001 DISP Connect the 5520A to the UUT as follows: 30.001 DISP [32] 5520A SENSE HI to UUT V-OHM 30.001 DISP [32] 5520A SENSE LO to UUT COM 30.001 DISP [32] 5520A NORMAL HI to UUT V-OHM 30.001 DISP [32] 5520A NORMAL LO to UUT COM 30.001 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 30.001 DISP the NORMAL leads to the SENSE leads AT the UUT. 30.002 5520 200.0nF 2.4U CW 31.001 5520 2.000uF 0.024U CW 32.001 5520 20.00uF 0.24U CW 33.001 5520 38.00uF 1.98U CW 34.001 HEAD {FREQUENCY TESTS} 34.002 DISP Connect the 5520A to the UUT as follows: 34.002 DISP [32] 5520A NORMAL HI to UUT V OHM 34.002 DISP [32] 5520A NORMAL LO to UUT COM 34.003 DISP Set the UUT to Hz ADp. 34.003 DISP 34.003 DISP Press the BLUE button to select FREQUENCY. 34.004 5520 100.0H 0.5U 40mV SI 2W 35.001 5520 1.000kH 0.005U 40mV SI 2W 36.001 5520 20.00kH 0.06U 40mV SI 2W 37.001 5520 200.0kH 1.4U 40mV SI 2W 38.001 5520 980kH 5.4U 400mV SI 2W 39.001 HEAD {ADp MODE} 39.002 DISP Press the BLUE button to select ADp. 39.003 5520 300.0mV S 2W 39.004 MATH S[1] = "UUT indicates between 2987 and 3013" 39.005 EVAL -s S[1] : Does UUT Indicate between 2987 and 3013? 40.001 ASK- R 40.002 HEAD {DC LOW CURRENT TESTS} 40.003 DISP Set the UUT to mA. 40.003 DISP 40.003 DISP Press the BLUE button to select DC. 40.004 DISP Connect the 5520A to the UUT as follows: 40.004 DISP [32] 5520A AUX HI to UUT mA 40.004 DISP [32] 5520A AUX LO to UUT COM 40.005 HEAD DC CURRENT TESTS: {4mA Range} 40.006 5520 4 2.000mA 0.010U 2W 41.001 HEAD DC CURRENT TESTS: {40mA Range} 41.002 5520 40 20.00mA 0.1U 2W 42.001 HEAD DC CURRENT TESTS: {400mA Range} 42.002 5520 400 200.0mA 1.0U 2W 43.001 HEAD {AC LOW CURRENT TESTS} 43.002 DISP Press the BLUE button to select AC. 43.003 HEAD AC CURRENT TESTS: {4mA Range} 43.004 5520 4 2.000mA 0.025U 50H SI 2W 44.001 5520 4 2.000mA 0.025U 1kH SI 2W 45.001 HEAD AC CURRENT TESTS: {40mA Range} 45.002 5520 40 20.00mA 0.25U 50H SI 2W 46.001 5520 40 20.00mA 0.25U 1kH SI 2W 47.001 HEAD AC CURRENT TESTS: {400mA Range} 47.002 5520 400 200.0mA 2.5U 50H SI 2W 48.001 5520 400 200.0mA 2.5U 1kH SI 2W 49.001 HEAD {AC HIGH CURRENT TESTS} 49.002 DISP Connect the 5520A to the UUT as follows: 49.002 DISP [32] 5520A 20A to UUT A 49.002 DISP [32] 5520A AUX LO to UUT COM 49.003 DISP Set the UUT function to A. 49.004 HEAD AC HIGH CURRENT TESTS: {10A Range} 49.005 5520 10 10.00A 0.15U 50H SI HC 2W 50.001 5520 10 10.00A 0.15U 1kH SI HC 2W 51.001 HEAD {DC HIGH CURRENT TESTS} 51.002 DISP Press the BLUE button to select DC. 51.003 HEAD DC HIGH CURRENT TESTS: {10A Range} 51.004 5520 10 10.00A 0.12U HC 2W 52.001 END